| Availability: | |
|---|---|
| Quantity: | |
|     PDF Export  |  |
HM-2000D
Product Description
Specification:
1. Standard parameters:ECG,RESP,NIBP,SPO2,PR/HR,TEMP.
2.13 types Arrhythmic Analysis, Multi-lead ECG waveforms display in Phase, 15 types Drug calculation and titration table.
3.Anti-ESU, anti-defibrillator; RA-LL impedance.
4.SPO2 can testing for 0.1% weak,high accuracy.
5.Unique breathing measurement technology, real-time monitoring, intelligent filter design, eliminate interference, measuring respiratory accurately.
6.Trend Coexist interface for all parameter to observe easily.
7.OxyCRG Dynamic View display, More intultive with all parameters.
8.Platform scalable with cable or wireless connecting to central monitoring.
9.Printer /Recorder is optional for real-time printing.
10.Brief operation interface, Touchscreen is optional.
Up to 2-5 hours working capacity of built-in rechargeable Li-ion battery.
11.12.1"high resolution color TFT LCD display.
12.Holographic waveform trends playback for 2 weeks and 12 hours off electricity saving.
13.Built-in networking to realize central monitoring function.
14.Optional: Touchscreen, VGA, 2-IBP, HDMI, ETCO2.
15.System start up only need 10 seconds with ARM-A5 Nuclear technology.
 
Standard accessory:
 
5-lead ECG cable                              1pc
Surface temp.probe                             1pc
    Adult NIBP cuff                             1pc
NIBP tube                                              1pc
Adult SpO2 sensor                             1pc
Power cord                                        1pc
Electrode                                      5pcs
User manual                                          1pc
 
 
 
Specifications
 
1.ECG
 
Lead Mode: 5 Leads or 3 Leads 
3 Leads:RA,LA,LL,Lead mode:I,II,III
5 Leads:RA,LA,LL,RL,V,Lead mode:I,II,III,aVR,aVL,aVF,V
Gain  ´250,´500,´1000,´2000
 
Heart Rate :
 
Range:    Adult:15 ~ 300bpm ;  Neonate / Pediatric:15 ~ 350 bpm
Accuracy:±1%or ±1bpm,
Resolution:1 bpm
Sensitivity  >200 uV
Differential Input Impedance >5MΩ
Bandwidth
Diagnostic:0.05~130Hz
Monitor:0.5~40Hz
Surgery:1~20Hz
 
 
2.RESP
 
Method: Impedance between R-F(RA-LL)
Scope of base impedance:200~4000Ω
Bandwidth: 0.1~ 2.5 Hz
 
RESP
Range: Adult:2~120BrPM   Neonate / Pediatric :2~150 BrPM
Resolution:1 BrPM
Accuracy:±2 BrPM
Suffocation warning:10~40s
 
3. NIBP
 
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring interval in AUTO MODE:1,2,3,4,5,10,15,30,60,90,120,180,240,480 MINS
Measuring interval in STAT MODE 5 MINS
PR range  40~240 bpm
Measuring Range and Resolution
Adult SYS  40~270mmHg
DIA   10~215mmHg
MEAN 20~235mmHg
PED  SYS  40~200mmHg
DIA   10~150mmHg
MEAN   20~165mmHg
NEO  SYS       40~135mmHg
DIA       10~100mmHg
MEAN   20~110mmHg
STATIC NIBP range: 0~300mmHg
Static NIBP accuracy:±3mmHg
Max Average tolerance ±5mmHg;Max. Standard tolerance: 8mmHg
Over-pressure protection:
Adult mode:  300 mmHg
Ped mode:         240 mmHg
Neo mode:  150 mmHg
 
 
4. SpO2
 
Measuring range:  0~100%
Resolution:          1%
Accuracy:           70~100%:±2;0%~69%
Pulse Rate:
Measuring range: 20~300bpm
Resolution:   1bpm
Accuracy:           ±3bpm
 
 
5.TEMP
 
Channels:     2   
Measuring:
Range:       0~50°C
Resolution:    0.1°C
Accuracy :     ±0.1°C(sensor tolerance no not included )
 
 
6. Power
 
Adapter 
Input: AC 90 V~240V, 50/60Hz
Output: DC 12V-20V
Built-in Battery:
TYPE:14.8V, 2200mAH Li-ion rechargeable battery
Operating time:3.5-4 hours working time under full power
Charging time: 2 hours
 
Label  ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART   0 ~ 300 mmHg,  PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP  -10 ~ 40 mmHg , P1/P2 -10 ~ 300 mmHg       Press Sensor
Sensitivity 5  uV/V/mmHg, Impedance 300-3000Ω
Resolution      1 mmHg
Accuracy      ±2% or ±1mmHg, which great
Actualization interval    about 1 Sec.
 
 Measuring and alarm range      0% ~ 13%
Resolution     1 mmHg
Accuracy     2 mmHg @  < 5.0% CO2(at ATPS)
RR                          3 - 150 bpm
 