HMA-3100
Healicom
Product Description
Features
• Automatic storage of test results, review and print all parameters in test results;
• Color touch screen, all parameters and histograms can be displayed in the same screen;
• Support the expansion of external devices such as mouse, keyboard and printer;
• Optical five reference ranges: newborn, children, adult women, adult men, and general use;
• Multiple clogging elimination methods: high-pressure burning jewel hole, high-pressure reverse rinsing and soaking;
• Automatic cleaning of sample probe;
• By VGA interface, external projector or external display screen can be connected;
• Chinese and English system language;
• Built-in thermal printer, serial port printing and other printing methods;
• With quality control management and calibration management function;
• Use the cyanide-free SFT method (environmentally friendly) to measure hemoglobin;
• Floating boundary algorithm and histogram abnormal prompt function.
Technical specifications | |
Test parameters | WBC 3-part differential counting, 23 parameters (including WBC, RBC, PLT color histogram); |
Measurement principle | counting by electrical impedance method, colorimetric method to measure HGB; |
Quality control method | L-J, automatic drawing and printing of quality control charts; |
Sample volume | Trace detection, 10μL peripheral blood or anticoagulant blood, pre-dilution mode 20μL; |
Precision | WBC (white blood cell) CV≤4.0%, RBC (red blood cell) CV≤2.0%, HGB (hemoglobin) CV≤2.0%, PLT (platelet) CV≤8.0%, MCV (average red blood cell volume) CV≤3.0% |
Accuracy | the allowable relative deviation range: WBC≤±15%, RBC≤±6.0%, HGB≤±6.0%, PLT≤±20.0%, HCT (hematocrit)≤±9.0% |
Blank count | WBC≤0.5×109/L, RBC≤0.05×1012/L, HGB≤2.0g/L, PLT≤10.0×109/L |
Carry-over | WBC≤3.5%, RBC≤2.0%, HGB≤2.0%, PLT≤5.0% |
Linearity deviation | WBC≤±5%, RBC≤±5%, HGB≤±3%, PLT≤±10% |
Related coefficient | WBC≥0.990, RBC≥0.990, HGB≥0.990, PLT≥0.990 |
Display | color LCD touch screen |
Detection channel | dual channel |
Test speed | 35 (or 60) specimens/hour, continuous working for 24 hours |
Data storage | It can automatically store more than 30,000 groups of complete results (each result has three histograms) |
Interface | RS232 interface, VGA interface |
Power supply | ~100V-240V; 50/60Hz |