Availability: | |
---|---|
Quantity: | |
PDF Export | |
FCA301
Healicom
Product Description
Feature:
● Random access fully automatic biochemistry analyzer
● New concept and more friendly software
● New techniques and spare parts are used for much better quality and function
● Automatic washing station for both sample and reagent probe
● Liquid level detection and collision protection for both sample and reagent probe
● Reliable cooling system for reagent chamber
● Powerful reaction cuvette washing system
● ISE module and barcode scanner (optional)
● Lower waste and pro environment
1. Sample/Reagent Probe
Sample/reagent probe with liquid sensor, anti-collision sensor and automatic washing function
2. Sample/Reagent Disk
40 positions (80 positions optional); 24-hour continuous refrigeration system with independent switch; 60
sample positions, available to use serum cup or blood tube
3. Cuvettes
High quality UV-transmitted plastic cuvettes, quartz gla ss cuvettes optional
4. 8 channels automatic washing system
Eight channels washing station for each cuvette, automatic record and deduct cuvette blank value
5. Independent mixer ensuring full reaction of sample/reagent position
Product Paramenters | |
Analytical System | Fully automatic, Random Access |
Method | End point, Rate, Fixed time, Immunoturbidimetric* 1-2 regent, Multistandard, reagent/serum blank, etc |
Assay Item | 80 assay |
Throughput | 1 Reagent: 300 tests/h (max) |
2 Reagent: 240 tests/h (max) | |
With ISE: 120 tests/h | |
Sample Position | 40 (or 80 optional) |
Including standard, control and STAT; either serum cup or primary tube available | |
Sample Volume | 1-100µL, 0.1µL/step |
Reagent Position | 2 * 40 |
Reagent Volume | R1: 1µL-400µL, 0.1µL/step R2: 1µL-400µL, 0.1µL/step |
Reagent/Sample Probe | With liquid level sensor, collision sensor, teflon coating and automatic washing |
Reaction Time | 0-999seconds |
Cuvette | High quality UV-transmitted plastic cuvette, or quartz glass cuvette on request |
Cuvette Time | 8 channels automatic washing system for reaction cuvettes |
Independent washing station for each sample/reagent probe and mixer | |
Alarm | Automatic alarm for malfuction (e.g: collision, reagent insufficiency, full waste container etc) |
Calibration | Linear, Non-linear, Multi point, K factor |
Sample Dilution/Retest | Sample can be diluted and retested if results are out of range or sample is insufficient. Also with calibrator auto diluting function |
Wavelength | 340-810nm |
Light Source | Halogen lamp |
Absorbance Resolution | 0.0001A |
Absorbance Accuracy | ±0.0003A(0~2.5A) |
Repeatability | ±2% |
Quality Control | Levy jennings quality control program at 3 levels |
Bar Code | Optional |
Printer | Multiple report formats available |
Language | English, French, Spanish, German, Russian, etc |
System | Windows XP or vistas; Windows 7 |
Power Supply | 220V/50Hz, or 110V/60Hz |
Dimension | 85 x 72 x 53cm |
Weight | 90Kg |