Power requirement | ≤400VA |
Temperature control mode | Air and skin servo control |
Air Temp Control Range | 20.0ºC-37.0ºC |
Temperature override mode to | 37.1ºC-39.0ºC |
Skin Temp Control Range | 20.0ºC-37.0ºC |
Temperature override mode to | 37.1ºC-38.0ºC |
Skin temp sensor accuracy | ≤0.3ºC |
Humidity display range | 5%-99% |
Temperature fluctuation | ≤0.5ºC |
Mattress temperature uniformity | ≤0.8ºC |
Temperature rise time | ≤45 min |
Noise level within hood environment | ≤55dB |
Alarms | Power failure alarm, Fan failure alarm, Sensor failure alarm |
High air temperature alarm | +3.0ºC |
Low air temperature alarm | -3.0ºC |
High skin temperature alarm | +1.0ºC |
Low skin temperature alarm | -1.0ºC |
Over air temperature alarm | ≤38.0ºC; ≤40.0ºC |
Over skin temperature alarm | ≤38.0ºC; ≤39.0ºC |
System failure alarm | |
Alarm silence function |