Flow(Optional) and pressure trigger | |
Air compressor (Optional) | |
Main Specification | |
Power | AC 220V±10%,(50Hz~60Hz)±2%, built-in battery for back-up |
Gas source | O2, Air |
Ventilation mode | VCV, SIMV, Sigh, SPONT, CPAP, PEEP, |
ST, S/T, Bi-level (Optional), Manual | |
Parameter monitoring: tidal volume, MV, breathing rate, inhaling time, Paw, PEEP, FiO2, Continuous pressure. | |
Waveform display: Pressure-time, Flow-time, Pressure-volume loop (optional) | |
Alarm | No tidal volume |
MV too high(low) | |
Paw too high(low) | |
FiO2 too high (low) | |
AC power failure | |
O2 or air failure | |
Breathing rate too high back-up battery too low |
SPECIFICATION
Physical specifications | |
Dimensions | 45*55*140cm |
Weight | 65kg |
Casters | Central brake with lock |
Screen | 10.2 inch |
Display | Color LCD |
Display graphics | Waveforms of P-T, F-T |
Loops | Optional: P-V, F-V |
Screen control | NEC, no dead spot at any angle |
Air compressor specifications | |
Power | AC 220V±10%, 50Hz±2% |
Output pressure | 0.4±0.1MPa |
Noise | <=65dB |
Ventilator Specifications | |
Modes of ventilation | |
VCV | Volume control ventilation |
SIGH | SIGH |
SIPPV(A/C) | Assistance Control |
SIMV | Synchronized Intermittent Mandatory Ventilation |
MANUAL | Manual |
CPAP | Continuous Positive Airway Pressure |
PSV | Pressure Support Ventilation |
PCV | Pressure Control Ventilation |
ARPV | N/A |
BIPAP | N/A |
Ventilator parameter | |
Tidal Volume range | 20-1500ml |
Atmosphere pressure | 960 hPa~1040 hPa |
Pressure range | 5-80 mbar |
Rispriation Rate Range | 1-99bpm (1-40bpm under SIMV) |
I:E Range | 4:1-1:8 |
Ptrigger Range | -20-20cmH2O |
Apnea Ti Range | 0-6s |
PEEP | |
Type | Integrated, Electronic control |
Range | 0-30cmH2O |
Alarm Limits | |
Visual | No tidal volume, MV,Paw, FiO2 limitation O2, Air failure, AC power failure, backup low |
Audio | No tidal volume, MV,Paw, FiO2 limitation |
Pipeline supply | |
Gas configuration | O2, Air |
Pipeline input range | 280-600kPa |
Battery Power | |
Battery Type | Lead-acid 12V, 4Ah |
Run-time | Typical 120 mins+ |
Pre-setting functions | |
Language | ENGLISH |
Calibration | AUTOMATIC |
General settings | N/A |
Applications | |
Patient therapy type | Audlt / Peadiatric Invasive |
Assistant therapy | N/A |
Environmental Conditions | |
Operating Temp | 5 ºC ~ 40 ºC |
Storage Temp | -10 ºC~+40 ºC |
Humidity | ≤80 % |
Optional | |
Air compressor |
AIR COMPRESSOR TECHNIQUE SPECIFICATIONS
Gas supply: O2, air 0.28~0.6MPa
Power supply: AC110/240V, 50Hz/60Hz, 65VA
Maximum security pressure: 8kPa
Compliance: 4mL/100Pa
Noise: <65dB(A)