| Technical specifications | |
| Infusion preset range (ml) | Feeding volume: 0-9999, adjusted by 1 Water feeding volume: 0-999, adjusted by 1 |
| Infusion(feeding or watering) accuracy | ±10% or±0.2ml/h(choose the larger value) |
| Feeding speed setting range (ml/h) | 1~800, adjusted by 1 |
| Feeding speed accuracy | ±10% or±0.2ml/h(choose the larger value) |
| Water feeding speed and accuracy (ml/h) | 1000±10% |
| Flushing speed and accuracy (ml/h) | 1000±10% |
| Single Flushing dose (ml) | 25±10% |
| Perfusion speed and precision (ml/h) | 1000±10% |
| Perfusion dose (ml) | 10ml±10% nutrition and 25ml±10% water |
| The Maximum infusion pressure | ≤0.1 MPa |
| The trigger alarm time at minimum flow rate | 42min18s |
| The trigger alarm time at medium flow rate | 1min30s |
| The amount of bolus produced by Medium speed up to the blocking alarm threshold | 0.49ml |
| Blocking alarm threshold (pressure) | 50KPa±10KPa |
| The recoverable time when alarm is automatically restored after the time is out | 1min50s - 2min |
| Pause time when alarm timeout | 1min50s - 2min |
| The heating temperature | When the feeding speed is greater than 20ml/h,15cm distance heater outlet is 25-38ºC |
| Bubble sensitivity | A bubble that can detect a single volume greater than 50uL |
| Intestinal feeder | Special intestinal feeding device |
| Alarm function | Tube occlusion, battery exhaustion, overtemperature |
| Prompt function | Tubing Bubble, Infusion Completion, Flush Completed, Battery Under voltage, Equipment Failure (ER Series), Pause Timeout, Network Power On / Off, Feeding /watering Switching |
| Noise | When operating at the maximum flow rate on the wooden desktop, noise level are below 50dB in the distance of 1 meter at the front, after and above. |
| Power supply and input | 100-240V ~ 50 /60Hz ≤40VA |
| Internal battery | Internal rechargeable lithium battery 11.1V; battery life > 8 hours |
| Output | ≤20VA |
| Rated output voltage | DC15V |
| Total Weight | About 1.5kg |
| Rate | Class II BF type continuous operation capacity type nutrition pump with internal power supply; Class IPX1 (anti-dripping device) |
| Dimensions | 135mm (L) ×105mm (W) ×160m (mH) |
| Accessory | 1 User's Manual and other document, 1 adapter, 1 set of intestinal feeding device and heater |