| Power requirement | ≤400VA |
| Temperature control mode | Air and skin servo control |
| Air Temp Control Range | 20.0ºC-37.0ºC |
| Temperature override mode to | 37.1ºC-39.0ºC |
| Skin Temp Control Range | 20.0ºC-37.0ºC |
| Temperature override mode to | 37.1ºC-38.0ºC |
| Skin temp sensor accuracy | ≤0.3ºC |
| Temperature fluctuation | ≤0.5ºC |
| Mattress temperature uniformity | ≤0.8ºC |
| Temperature rise time | ≤45 min |
| Noise level within hood environment | ≤55dB |
| Alarms | Power failure alarm, Fan failure alarm, Sensor failure alarm |
| High air temperature alarm | +3.0ºC |
| Low air temperature alarm | -3.0ºC |
| High skin temperature alarm | +1.0ºC |
| Low skin temperature alarm | -1.0ºC |
| Over air temperature alarm | ≤38.0ºC; ≤40.0ºC |
| Over skin temperature alarm | ≤38.0ºC; ≤39.0ºC |
| System failure alarm | |
| Alarm silence function | |