Feature:
● Random access fully automatic biochemistry analyzer
● New concept and more friendly software
● New techniques and spare parts are used for much better quality and function
● Automatic washing station for both sample and reagent probe
● Liquid level detection and collision protection for both sample and reagent probe
● Reliable cooling system for reagent chamber
● Powerful reaction cuvette washing system
● ISE module and barcode scanner (optional)
● Lower waste and pro environment
1. Sample/Reagent Probe
Sample/reagent probe with liquid sensor, anti-collision sensor and automatic washing function
2. Sample/Reagent Disk
40 positions (80 positions optional); 24-hour continuous refrigeration system with independent switch; 60
sample positions, available to use serum cup or blood tube
3. Cuvettes
High quality UV-transmitted plastic cuvettes, quartz gla ss cuvettes optional
4. 8 channels automatic washing system
Eight channels washing station for each cuvette, automatic record and deduct cuvette blank value
5. Independent mixer ensuring full reaction of sample/reagent position
| Product Paramenters | |
| Analytical System | Fully automatic, Random Access |
| Method | End point, Rate, Fixed time, Immunoturbidimetric* 1-2 regent, Multistandard, reagent/serum blank, etc |
| Assay Item | 80 assay |
| Throughput | 1 Reagent: 300 tests/h (max) |
| 2 Reagent: 240 tests/h (max) | |
| With ISE: 120 tests/h | |
| Sample Position | 40 (or 80 optional) |
| Including standard, control and STAT; either serum cup or primary tube available | |
| Sample Volume | 1-100µL, 0.1µL/step |
| Reagent Position | 2 * 40 |
| Reagent Volume | R1: 1µL-400µL, 0.1µL/step R2: 1µL-400µL, 0.1µL/step |
| Reagent/Sample Probe | With liquid level sensor, collision sensor, teflon coating and automatic washing |
| Reaction Time | 0-999seconds |
| Cuvette | High quality UV-transmitted plastic cuvette, or quartz glass cuvette on request |
| Cuvette Time | 8 channels automatic washing system for reaction cuvettes |
| Independent washing station for each sample/reagent probe and mixer | |
| Alarm | Automatic alarm for malfuction (e.g: collision, reagent insufficiency, full waste container etc) |
| Calibration | Linear, Non-linear, Multi point, K factor |
| Sample Dilution/Retest | Sample can be diluted and retested if results are out of range or sample is insufficient. Also with calibrator auto diluting function |
| Wavelength | 340-810nm |
| Light Source | Halogen lamp |
| Absorbance Resolution | 0.0001A |
| Absorbance Accuracy | ±0.0003A(0~2.5A) |
| Repeatability | ±2% |
| Quality Control | Levy jennings quality control program at 3 levels |
| Bar Code | Optional |
| Printer | Multiple report formats available |
| Language | English, French, Spanish, German, Russian, etc |
| System | Windows XP or vistas; Windows 7 |
| Power Supply | 220V/50Hz, or 110V/60Hz |
| Dimension | 85 x 72 x 53cm |
| Weight | 90Kg |